9 mm
Resolution
static 1)
16 100 nm
dynamic 2)
48 192 480 nm
Linearity 3)
Displacement and distance ±0.9 1.5 IFS2402-1,5 IFS2402-4
Measuring range 0. optical fiber)
1)
Average from 512 values kHz, the mid the measuring range onto optical flat
2)
RMS noise relates mid measuring range kHz)
3)
All data constant ambient temperature (25±1 °C).
5)
No thickness measurement possible Distance measurement only possible thickness glass measuring range. The accuracy decreases when approaching the limit values. Measurements metal only possible limited extent.
MR Measuring range
SMR Start measuring range
Dimensions mm, not scale
Confocal chromatic miniature sensors
confocalDT IFS2402
Miniature sensors with
axial radial beam path
Submicron resolution
For precise distance
measurements
Small light spot
68
15
ø4
fiber
optic
ø2.16
Model IFS2402-0.10
Target material reflective, diffuse well transparent surfaces (e.g.5 1.7 0.2 ±1. Measurement plane-parallel test glass. glass) 5)
Connection
integrated optical fiber with E2000/APC connector; extension m;
bending radius: static mm; dynamic mm
Installation Clamping (mounting adapter see accessories)
Temperature range
Storage -20 +70 °C
Operation +70 °C
Shock (DIN 60068-2-27) 15g axis, 1000 shocks each
Vibration (DIN 60068-2-6) 500 axis, cycles each
Protection class (DIN 60529) IP64 (front)
Material Stainless steel housing, glass lenses
Weight approx.
4)
Maximum measuring angle the sensor that produces usable signal reflecting surfaces. 1.40 0. Acceptance report enclosed with delivery.20 0. measuring angle 4)
±18° ±5° ±3°
Numerical aperture (NA) 0.5 mm
Start measuring range approx.5 3.2 µm
Light spot diameter µm
Max. 186 (incl.1
SMR
MR
Kink
protection
and
strain
relief