Configuration the calibrator
Calibrator should tested directly from the front panel terminals. Connect calibrator output standard multimeter type according the table Tab select function of
ohmmeter.
2.Verification procedure
MEATEST.1 %
* Multifunction calibrator Meatest M-140, M142, Fluke 5500A similar with voltage range 1000
V.1%.r.
* Process calibrator Burster 4423, Fluke 745A similar with current range and accuracy
better than .2%. Performance verification may performed
after warm-up period i. Connect the calibrator the mains and let them switched for least one 15minutes laboratory at
23±1 oC.
R tio n
1. minutes after switching on. Set resistance values 10kQ 100 the calibrator.
Basic steps oftheperformance verification test
Verification procedure consists following steps:
• range from 1TQ
• lta ranges kV
• range mA
• test points 10n, 50n, 100nF
Procedure
Following part describes procedure performance verification test.
* High voltage source Heinzinger PNC-10000 similar with output voltage accuracy 0.
* 81/2 digit multimeter Agilent 3458A pA-meter Keithley 2635A similar with current range100
nA mA
* RLC meter Agilent 4263A, Agilent 4278A, other with accuracy 0.___________________________ M191 Insulation Tester Calibrator
Procedure recommended for verifying parameters the calibrator described this chapter.e.o.
User Manual v16 33
. Calibrator have temperature stabilize condition
at minimum hours before performance verification test started. Check reading the standard multimeter and
compare with limits the table TabV.
Required equipment
Following instruments are required for calibration:
* 81/2 digit multimeter type Fluke 8508A similar with 0. Recommended test points including applied
method specification and allowed limits are shown table Tab V.01 accuracy and range GQ. s.
3