Both complete and partial calibration possible.
Calibration process can interrupted left any time pushing the button SETUP. Confirm with ENTER button rotary button.
Go next requested calibration point. Calibrator returns the
last selected function. Output terminals are automatically switched on
at the same time ready for measuring the selected calibration point.
Input resistance internal mA-meter approx.
Note: When resistance points are calibrated two-terminals way connection the calibrator tostandard
ohmmetershouldbeapplied.
• When built mA-meter calibrated, connect current requested valuefrom standard unit and
confirmwith ENTER button rotary button save new calibration data.
List calibration points includes:
• Single resistance segments summary range from 1TQ, R00 R45
• Single capacitors 10, 50, 100 nF, C00 C02
• voltage 5000 V00 and V01
• current mA, I00 and I01
28 User Manual
.
• When built kV-meter calibrated, connect voltage requested value fromstandard unit and
confirmwith ENTER button rotary button save new calibration data. Standard current source with compliance
voltagemin. New calibration value saved immediately when
ENTER button rotary button pushed for new value confirmation.r. s.
Selection and setting ofcalibration point
After the calibration menu displayed, one partial calibrations can selected.
Calibration points
Each function the calibrator has assigned fixed calibration points which have set during the calibration.o.
• Connect RCL standard meter when capacitance calibration point beingcalibrated andwrite down
new resistance value. Confirm with ENTER button rotary button. When only some calibration points are changed, all other non
changed calibration points are stored with original data. Use rotary button move the
cursor through the list.M191 Insulation Tester Calibrator MEATEST. 2.
When change the calibration value requested push the rotary button ENTER button once more. Having selected the required function calibrated, press the rotary button ENTER
button enable the calibration value right column for editing. The
original calibration value refreshed. 15Vhas tobeapplied.5 kQ.
• Measure the resistance value using standard ohmmeter when resistance segment calibrated and write
down new resistance value