The
original calibration value refreshed.
• When built kV-meter calibrated, connect voltage requested value fromstandard unit and
confirmwith ENTER button rotary button save new calibration data. Confirm with ENTER button rotary button.
Note: When resistance points are calibrated two-terminals way connection the calibrator tostandard
ohmmetershouldbeapplied. Calibrator returns the
last selected function. 15Vhas tobeapplied.
Calibration points
Each function the calibrator has assigned fixed calibration points which have set during the calibration. When only some calibration points are changed, all other non
changed calibration points are stored with original data. s.
• Measure the resistance value using standard ohmmeter when resistance segment calibrated and write
down new resistance value.
Selection and setting ofcalibration point
After the calibration menu displayed, one partial calibrations can selected.
Go next requested calibration point.
List calibration points includes:
• Single resistance segments summary range from 1TQ, R00 R45
• Single capacitors 10, 50, 100 nF, C00 C02
• voltage 5000 V00 and V01
• current mA, I00 and I01
28 User Manual
. New calibration value saved immediately when
ENTER button rotary button pushed for new value confirmation.M191 Insulation Tester Calibrator MEATEST. Use rotary button move the
cursor through the list.
Calibration process can interrupted left any time pushing the button SETUP.
Input resistance internal mA-meter approx.
• When built mA-meter calibrated, connect current requested valuefrom standard unit and
confirmwith ENTER button rotary button save new calibration data. Standard current source with compliance
voltagemin.
Both complete and partial calibration possible.5 kQ. Output terminals are automatically switched on
at the same time ready for measuring the selected calibration point. Having selected the required function calibrated, press the rotary button ENTER
button enable the calibration value right column for editing.
• Connect RCL standard meter when capacitance calibration point beingcalibrated andwrite down
new resistance value. Confirm with ENTER button rotary button.r.o.
When change the calibration value requested push the rotary button ENTER button once more. 2