Poznámky redaktora
. Partial resistors
of the internal serial resistance decade and calibration points internal meters can readjusted during the
calibration predefined order.e. PIB address. Calibration mode
The M191 calibrator includes calibration procedure, which allows calibration the calibrator.
11. The parameter cannot changed..o..
M191 Insulation Tester Calibrator________________________________________ MEATEST, s.
9. The address set the manufacturer. The calibration can controlled using the buttons and menu the calibrator.pressing ENTER.e. CPU firm are x
CPU block firmware version. UP, DOWN display buttons can used select any valid
GBIP address the range 30. UP/D N
Indicates the communication speed RS232 bus. all functions are calibrated all recommended points
• partially, i. only selected functions are calibrated selected points
26 User Manual
. UP/DOWN display buttons can used select 9600, 19200,
28800, 38400, 57600, 115200.
Calibration principles
The calibrator can calibrated:
• completely, i. Non-zero calibration code not displayed further the display. Perfect communication with the requires equal values set the and the
calibrator... PIB/RS232
Displays the type interface used control the calibrator from PC. pressing GPIB RS232 buttons, the
respective type can selected. firm are x
LVR block firmware version.. N
Displays the calibrator’s address the GPIB bus.
6 Serial num ber x
Displays the serial number ofthe calibrator.
10.
8 S232 baud . Active terfa ce. non-zero calibration code set, correct calibration code must entered access the
calibration mode.
7.r.... The calibrator can remotely controlled only using the selected interface.
6.e. The parameter cannot changed. only selected functions are calibrated all recommended points
• partially, i. The parameter cannot changed