Poznámky redaktora
r. Calibration mode
The M191 calibrator includes calibration procedure, which allows calibration the calibrator.
10. only selected functions are calibrated all recommended points
• partially, i.. The parameter cannot changed. Non-zero calibration code not displayed further the display. Active terfa ce.pressing ENTER. The calibration can controlled using the buttons and menu the calibrator.. N
Displays the calibrator’s address the GPIB bus.. The parameter cannot changed.. The address set the manufacturer.
11.
M191 Insulation Tester Calibrator________________________________________ MEATEST, s. UP/D N
Indicates the communication speed RS232 bus.o. The calibrator can remotely controlled only using the selected interface.
Calibration principles
The calibrator can calibrated:
• completely, i. non-zero calibration code set, correct calibration code must entered access the
calibration mode. Partial resistors
of the internal serial resistance decade and calibration points internal meters can readjusted during the
calibration predefined order. only selected functions are calibrated selected points
26 User Manual
.e. firm are x
LVR block firmware version.
6.. PIB/RS232
Displays the type interface used control the calibrator from PC. PIB address.e. CPU firm are x
CPU block firmware version.
6 Serial num ber x
Displays the serial number ofthe calibrator. The parameter cannot changed. UP, DOWN display buttons can used select any valid
GBIP address the range 30.. Perfect communication with the requires equal values set the and the
calibrator. all functions are calibrated all recommended points
• partially, i. UP/DOWN display buttons can used select 9600, 19200,
28800, 38400, 57600, 115200.e... pressing GPIB RS232 buttons, the
respective type can selected.
7..
8 S232 baud .
9