Poznámky redaktora
The parameter cannot changed. only selected functions are calibrated selected points
26 User Manual
.o.
8 S232 baud .e. Calibration mode
The M191 calibrator includes calibration procedure, which allows calibration the calibrator.
9.
6 Serial num ber x
Displays the serial number ofthe calibrator. The parameter cannot changed.. The parameter cannot changed.
M191 Insulation Tester Calibrator________________________________________ MEATEST, s. The calibration can controlled using the buttons and menu the calibrator.
11. N
Displays the calibrator’s address the GPIB bus. Non-zero calibration code not displayed further the display. PIB/RS232
Displays the type interface used control the calibrator from PC.. all functions are calibrated all recommended points
• partially, i. firm are x
LVR block firmware version. UP/D N
Indicates the communication speed RS232 bus... The calibrator can remotely controlled only using the selected interface..
10.
Calibration principles
The calibrator can calibrated:
• completely, i. CPU firm are x
CPU block firmware version. Perfect communication with the requires equal values set the and the
calibrator. The address set the manufacturer. Active terfa ce.. UP/DOWN display buttons can used select 9600, 19200,
28800, 38400, 57600, 115200.pressing ENTER.e..r.e. UP, DOWN display buttons can used select any valid
GBIP address the range 30.
6.. Partial resistors
of the internal serial resistance decade and calibration points internal meters can readjusted during the
calibration predefined order. non-zero calibration code set, correct calibration code must entered access the
calibration mode.
7. pressing GPIB RS232 buttons, the
respective type can selected. PIB address. only selected functions are calibrated all recommended points
• partially, i.