Poznámky redaktora
only selected functions are calibrated selected points
26 User Manual
. Active terfa ce.
9..
11.
10. The parameter cannot changed. PIB address.
8 S232 baud . UP, DOWN display buttons can used select any valid
GBIP address the range 30. CPU firm are x
CPU block firmware version.e.
M191 Insulation Tester Calibrator________________________________________ MEATEST, s.
6 Serial num ber x
Displays the serial number ofthe calibrator. only selected functions are calibrated all recommended points
• partially, i. firm are x
LVR block firmware version. The calibrator can remotely controlled only using the selected interface.... Calibration mode
The M191 calibrator includes calibration procedure, which allows calibration the calibrator.e. Partial resistors
of the internal serial resistance decade and calibration points internal meters can readjusted during the
calibration predefined order.. non-zero calibration code set, correct calibration code must entered access the
calibration mode. UP/D N
Indicates the communication speed RS232 bus..
7.pressing ENTER. N
Displays the calibrator’s address the GPIB bus.
Calibration principles
The calibrator can calibrated:
• completely, i. pressing GPIB RS232 buttons, the
respective type can selected. Perfect communication with the requires equal values set the and the
calibrator.r.e. Non-zero calibration code not displayed further the display.. The calibration can controlled using the buttons and menu the calibrator.. The parameter cannot changed. PIB/RS232
Displays the type interface used control the calibrator from PC. The address set the manufacturer.o.
6. all functions are calibrated all recommended points
• partially, i. UP/DOWN display buttons can used select 9600, 19200,
28800, 38400, 57600, 115200. The parameter cannot changed.