Poznámky redaktora
only selected functions are calibrated selected points
26 User Manual
..
9.e.
6.
7. Perfect communication with the requires equal values set the and the
calibrator. The parameter cannot changed.
8 S232 baud .. firm are x
LVR block firmware version. Active terfa ce.
10.. The address set the manufacturer.
6 Serial num ber x
Displays the serial number ofthe calibrator. UP, DOWN display buttons can used select any valid
GBIP address the range 30.e. The parameter cannot changed..e.pressing ENTER.r. PIB/RS232
Displays the type interface used control the calibrator from PC.
11... non-zero calibration code set, correct calibration code must entered access the
calibration mode.
M191 Insulation Tester Calibrator________________________________________ MEATEST, s. only selected functions are calibrated all recommended points
• partially, i. UP/D N
Indicates the communication speed RS232 bus..o.. CPU firm are x
CPU block firmware version. The calibrator can remotely controlled only using the selected interface. Calibration mode
The M191 calibrator includes calibration procedure, which allows calibration the calibrator. Non-zero calibration code not displayed further the display. The parameter cannot changed. all functions are calibrated all recommended points
• partially, i.. pressing GPIB RS232 buttons, the
respective type can selected.
Calibration principles
The calibrator can calibrated:
• completely, i. PIB address. UP/DOWN display buttons can used select 9600, 19200,
28800, 38400, 57600, 115200. Partial resistors
of the internal serial resistance decade and calibration points internal meters can readjusted during the
calibration predefined order. The calibration can controlled using the buttons and menu the calibrator. N
Displays the calibrator’s address the GPIB bus