Poznámky redaktora
When presence the test
voltage recognized, calibrator goes over RUNNING mode.t.
RUNNING Calibrator simulation sequence has been launched.
- Total time.o.0 Mil' cDfl: 1.
- max parameter maximum applicable test voltage.11 ProgrammingtheDPPmode
Following parameters can entered:
- PARAMETER field.
D. Calibrator switches the setup resistances to
the output terminals preset time intervals.
B.
- DAR/PR/PI coefficient. this mode
is waiting for test voltage from UUT start simulation sequence. Use only test leads with guaranteed isolation according test voltage
range UUT. Select requested type of
parameter and confirm with ENTER button rotary button. The test voltage must higher than
100 VDC start the RUNNING sequence. Use rotary button select and enter parameter R0, PARAMETER and coefficient.
During running mode currently applied test voltage and total elapsed time are displayed.
STANDBY When OUTPUT button pushed, calibrator goes over STANDBY mode. Connect UUT the output terminals.
C. From the entered coefficient and nominal resistance
calibrator calculates resistance level which will connected during RUNNING mode the output
terminals
Following additional parameters are displayed DAR mode.MEATEST, s.9. M191 Insulation Tester Calibrator
OFF OFF mode the output terminals are disconnected. actual resistance value which currently connected the RUNNING mode the
output terminals.
Range the value from 0..000
pnn.
User Manual v16 21
.: ---------
Umax" kU
OFF f
Local
Gnd
Te^t 1ta*ie:-U.r. When the sequence finished, calibrator stays in
RUNNING mode until the output terminals are disconnected using OUTPUT ON/OFF button.
- out value.0 s
Mode :
DPP
Accuracy
Fig.
- parameter. This the resistance which connected the output terminals STANDBY mode.
- Test voltage. Requested value the coefficient which asked simulate can entered.5 99. Both type requested parameter,
nominal resistance and parameter index can set. The time which has been elapsed after the starting RUNNING sequence. When requested parameters are set up, push the OUTPUT button. Select DPP mode pushing repeatedly FUNCTION FUNCTION DOWN button until DPP label
appears the display.
In DPP mode the display shows information follows (OFF submode):
R0S 100. kU
Total time: 0.
The parameter range from 100 GQ. Write requested resistance value using the
numerical keypad and range buttons kQ, MQ, GQ, and push ENTER button rotary button confirm. Calibrator connects resistance value
R0 the output terminals and waiting for test voltage from UUT.
Operating procedure
A.
- Accuracy actually connected resistance %. One defined parameters DAR/PR/PI can selected. Reading currently applied test voltage displayed. Use test leads delivered with UUT whenever possible.
Write the coefficient and confirm pushing the ENTER button