The qualification must comprise professional
education, knowledge and experience well knowledge the applicable standards and local
provisions.3 Test acco 0751-1
The DIN VDE 0751-1 standard applies for tests medical electrical devices, medical electrical
systems parts those devices systems complying with the DIN 60601-1 (VDE0 750 Part
1) standard which are carried out prior commissioning, inspection, maintenance during periodic
testing order evaluate the safety those devices systems their parts.
(taken from VDE 0751-1: issue 2008-08, section 4. Insulation measurement must not carried out,
if excluded the accompanying documents according the manufacturer's specifications.
(taken from VDE 0751-1:2001-10, section 1.
In the current issue the DIN VDE 0751-1 standard, the measurement the insulating resistance
is required, considered convenient.1)
Prior testing, accompanying documents have checked for the manufacturer's recommenda
tions regarding maintenance and repair, including conditions and precautionary measures. Personnel evaluating safety must able detect possible consequences and dangers
that might caused devices not complying with the requirements.
Lim iting valu for the insulating resistance are not specified the DIN VDE 0751-1 standard
(issue 2008). For devices which
are not built according DIN 60601-1 (VDE 0751 Part 1), this standard can applied anal
ogously under consideration the safety standards applying the manufacturing the device. The DIN VDE 0751-1 standard recommends the following order the tests to
be carried out:
A cco DIN 0751-1 llo seq tests reco ended :
I
Test selection
I
Visual inspection
I
Protective conductor
resistance
I
Leakage currents
I
(Insulating resistance)
Functional test
I
Report results
I
Evaluation
Inspection and preparation for
intended use
08/ 2012 58
.1)
The tests must carried out qualified personnel. Here, possible refer recommendations the manufacturer previously
measured values.BENNING BENNING 750 A