Poznámky redaktora
.pressing ENTER.
10. non-zero calibration code set, correct calibration code must entered access the
calibration mode. CPU firm are x
CPU block firmware version. UP, DOWN display buttons can used select any valid
GBIP address the range 30...
M191 Insulation Tester Calibrator________________________________________ MEATEST, s.
6. UP/DOWN display buttons can used select 9600, 19200,
28800, 38400, 57600, 115200. PIB/RS232
Displays the type interface used control the calibrator from PC.
8 S232 baud .
7. Partial resistors
of the internal serial resistance decade and calibration points internal meters can readjusted during the
calibration predefined order.... The calibration can controlled using the buttons and menu the calibrator. The calibrator can remotely controlled only using the selected interface. N
Displays the calibrator’s address the GPIB bus. only selected functions are calibrated selected points
26 User Manual
.e. Non-zero calibration code not displayed further the display.
Calibration principles
The calibrator can calibrated:
• completely, i.r.e. UP/D N
Indicates the communication speed RS232 bus.
9. firm are x
LVR block firmware version.o. The parameter cannot changed. The parameter cannot changed. PIB address.. only selected functions are calibrated all recommended points
• partially, i. all functions are calibrated all recommended points
• partially, i. The parameter cannot changed. The address set the manufacturer. Active terfa ce..e. Perfect communication with the requires equal values set the and the
calibrator..
11.
6 Serial num ber x
Displays the serial number ofthe calibrator. pressing GPIB RS232 buttons, the
respective type can selected. Calibration mode
The M191 calibrator includes calibration procedure, which allows calibration the calibrator